Product Description
— Universal device for all infusions and transfusions
— Accurate &Work with all brands of infusion tube
Accuracy Error ≤ ±5% after calibrated , work with all brands of IV tube: 10d/ml 、15d/ml 、20d/ml 、60d/ml (3.4mm – 4.5mm)
— Safety and Reliable
2 CPU systems for self-detection;
Multiple sensors can detect and monitor the infusion status
— Multi-functions
12 infusion mode including Programmable infusion mode;
Drug Library and history records data can be transferred;
WIFI with communication Interface for information system;
Warmer for heating.
Specifications
Model | Single-channel Infusion pump WIT-601A | |
Communication ( Optional) | WIFI, with communication protocol Interface for information system | |
Warmer (Optional) | Warmer on the side of pump to keep the medical liquid warm | |
Drug Library | Library size configurable: More than 1000 drugs &54 drug dose unit | |
History Records | Up to 20,000 infusion records | |
Data Transfer | RS485 port(compatible with RS232), Wired interface connectable with PC to upload& download data . | |
Display | High brightness color LCD 3.2inch screen , Adjustable brightness. | |
Infusion Mode | Rate/VTBI 、Time/VTBI 、Rate/Time 、Volume 、Drop 、Micro 、Simple Rate、 Programmable 、DOSE Weight 、Drug lib 、Auto Bolus 、Manual Bolus. | |
Programmable | 8 programmable sequences of ‘ Time & Flow rate & VTBI’ | |
IV SET | Open System: 10d/ml 、15d/ml 、20d/ml 、60d/ml IV tube can be used for all brands of I V TUBE; More than 50 brands IV tube feature data can be stored inside the pump. | |
Closed System: Can be customized | ||
Accuracy Error | Volume Mode :≤ ±5% | Drop Mode :≤ ±2% |
Calibration | Automatic or manual calibration; | |
Flow Rate | Volume:0. 10ml/h~1200ml/h | Drop:1d/min~400d/min |
Flow rate increment | 0.01ml/h | |
VTBI | 0. 1~9999.9 ml | |
Total Volume | 0.1~9999.9ml | |
KVO | Default:1ml/h ; & 1ml/h~20ml/h can be set. | |
Purge | 600ml/h | |
Bolus | Keep pushing down ‘Bolus’ button to run the default Bolus parameters. Or click the ‘Bolus’ button to enter the Bolus parameter editable interface. | |
Upstream pressure | High / Middle / Low, or OFF | |
Downstream pressure | 20 pressure Levels: 200mmHg – 1200mmHg ( mmHg & kPa ) | |
Bubble | Detected by ultrasonic sensor with adjustable 5 levels, 50μL ~ 800μL, | |
Drop Sensor | Infrared sensor to detect the drops, monitoring the infusion status | |
Alarms | Audio and Visual alarm for : Bubble, Door open, Finish, Pause timeout, Down Occlusion, Up Occlusion, Lower Battery or damage, Motor error, Incorrect IV Set-type, Pressure test failure, Motor failure, Abnormal drop rate, Heater failure and etc. 30 kinds safety monitoring alarm | |
Self -detection | 2 CPU systems; Self-detection at startup and running time | |
Non-interrupt | Parameters can be modified when pump running | |
KeypadLock | Automatically or Manually lock the keyboard | |
AC Power | AC100V ~ 240V ,50Hz/60Hz ,25VA; &Vehicle-mounted power supply adapter (Optional ) | |
Battery | NI-MH battery,DC9.6V, 2000mAh ( 6 hours at 25ml/h rate ) | |
Electrical Safety Class | Class I, Type CF, IPX4 | |
Install | Can be used either on Iv stand, rails or as a table top device | |
Weight & Dimension | 2.0Kg; & 130mm×120mm×195mm | |
Working Environment | 5℃ ~ 40℃ ; 20% ~ 90% humidity; 70kPa ~ 106kpa atmospheric | |
Application | Medical locations such as intensive care unit, pediatric ward, obstetrics and gynecology department for constant speed intravenous infusion of drugs | |