Product Description
— Accurate &Work with all brands of syringe
Accuracy Error ≤ ±2% after calibrated , work with all brands
of 5ml 、10ml、20ml、30ml、50ml(60ml) syringes which can be automatically sensed
— Safety and Reliable
2 CPU systems for self-detection;Multiple sensors can detect and monitor the injection status
— Multi-functions
10 injection mode including Programmable injection mode;
Drug Library and history records data can be transferred;
WIFI with communication Interface for information system
Specifications
Model | Single-channel syringe pump WIT-301A |
Communication(Optional) | WIFI, with communication protocol Interface for information system |
Drug Library | Library size configurable: More than 1000 drugs & 54 drug dose unit; |
History Records | Up to 20,000 infusion records |
Data Transfer | RS485 port(compatible with RS232), Wired interface connectable with PC to upload& download data . |
Display | High brightness color LCD 3.2inch screen , Adjustable brightness. |
Injection Mode | Rate/VTBI、Time/VTBI、Rate/Time、Micro、Simple Rate、 Programmable、 DOSE Weight、 Drug Lib 、Auto Bolus 、Manual Bolus. |
Programmable | 8 programmable sequences of ‘ Time & Flow rate & VTBI’ |
Syringe | 5ml 、10ml、20ml、30ml、50ml(60ml) syringes can be used and automatically identified Open system for all brands of syringes More than 50 brands syringes feature data can be stored inside the pump. |
Accuracy Error | ≤ ±2% |
Calibration | Automatic or manual calibration ; |
Flow Rate | 5ml syringe: 0.10- 150 ml/h 10ml syringe: 0.10-300 ml/h 20ml syringe: 0.10-400 ml/h 30ml syringe: 0.10-600 ml/h 50ml (60 ml ) syringe: 0.10- 1200ml/h |
Flow rate increment | 0.01ml/h |
VTBI | 0. 1~9999.9 ml |
Total Volume | 0.1~9999.9ml |
KVO | Default:1ml/h ; & 0.1ml/h~20ml/h can be set. |
Bolus | Keep pushing down ‘Bolus’ button to run the default Bolus parameters. Or click the ‘Bolus’ button to enter the Bolus parameter editable interface. |
Anti-Bolus | Automatically back off to release the pressure of occlusion to avoid risk of ‘Bolus’ . |
Occlusion Pressure | 20 pressure Levels: 200mmHg – 1200mmHg ( mmHg & kPa) |
Alarms | Audio and Visual alarm for : Occlusion, Occlusion drop rapidly, Near Finish, Finish, KVO, Pause timeout, Pusher error, Compression bar error, Low Battery or Battery Discharged, Motor error, Parameter error. Interrupt of AC Power etc. |
DERS | Automatically check and adjust to eliminate the mechanical gap to reduce dose error at startup |
Self -detection | 2 CPU systems; Self-detection at startup and running time |
Non-interrupt | Parameters can be modified when pump running |
KeypadLock | Automatically or Manually lock the keyboard |
AC Power | AC100V ~ 240V ,50Hz/60Hz ,25VA; &Vehicle-mounted power supply adapter (Optional) |
Battery | NI-MH battery, DC9.6V, 2000mAh ( 7 hours at 5ml/h rate ) |
Electrical Safety Class | Class I, Type CF, IPX4 |
Install | Can be used either on Iv stand, rails or as a table top device; Multiple pumps can be superimposed up and down |
Weight &Dimension | 1.9Kg; & 140mm×145mm×325mm |
Working Environment | 5℃ ~ 40℃ ; 20% ~ 90% humidity; 70kPa ~ 106kpa atmospheric |
Application | Medical locations such as intensive care unit, pediatric ward, obstetrics and gynecology department for constant speed intravenous infusion of drugs |